App Notes & Whitepapers

Optimizing an ALH with Custom Glycerol Solutions Using MVS® and ArtelWare® Version 2.3

Optimizing an Automated Liquid Handler with Custom Glycerol Solutions Using the MVS® with ArtelWare® Software

The Multichannel Verification System (MVS) in conjunction with ArtelWare version 2.3 was employed to prepare MVS Custom Solutions, which are specialized test solutions chemically distinct from the commercially available MVS QualAssure Solutions.

These Custom Solutions enable testing volumes ranging from approximately 0.4–49.9 μL in 96-well plates and approximately 0.1–9.9 μL in 384-well plates. Custom Solutions with glycerol concentrations of 17%, 46%, and 68% v/v in water were prepared by combining glycerin, deionized water, and MVS Stock Solution 2, utilizing the Custom Solution liquid handler verification feature in ArtelWare.

These solutions were subsequently utilized to fine-tune liquid classes for 10 μL dispenses on an automated liquid handler (ALH) using the Dual-Dye Ratiometric Photometric Method provided by the MVS system.

Download App Note

Please fill out this brief form to download this content.

One moment while the form loads...

Please note, browsing in private or incognito may not be supported and may not allow forms and other elements to load correctly.

Related Resources

Learn more by checking out our additional educational resources below.

Liquid Handling Case Study

Keeping Plasma Safe: Haematologic Technologies Inc. Implements the Artel PCS for Quality Assurance

View

Liquid Handling Case Study

Flying on Course: FAA’s Forensic Toxicology Lab Enhances Quality and Compliance with the Artel PCS

View

Liquid Handling Case Study

The Impact of the Artel MVS: When Analyzing Cancer at the Molecular Level, Microliters Matter

View

Liquid Handling App Note

Evaluating Volumetric Performance of Beckman Echo Acoustic Liquid Handlers with Artel MVS vs. Fluorescence Test Methods

View

Liquid Handling App Note

Optimizing an ALH with Custom Glycerol Solutions Using MVS® and ArtelWare® Version 2.3

View

Liquid Handling App Note

Calibration of Multichannel Pipettes Using the Artel MVS®

View

Liquid Handling Webinar

Virtual Demo | Artel MVS® Multichannel Verification System

View

Liquid Handling Webinar

Virtual Demo | Artel PCS® Pipette Calibration System

View

Liquid Handling Webinar

Assuring Microliter Volume Handling in Molecular Analysis Using Dual-dye Ratiometric Photometry

View

Liquid Handling Posters

Artel Poster: 10 Tips To Improve Your Pipetting Technique

View

Liquid Handling Posters

8 Easy Steps for Performing Pipette Preventive Maintenance

View

Liquid Handling Posters

Effect of Liquid Handling Quality Control on Biological Assay Performance

View

Liquid Handling Knowledge Center

Metrological Traceability of In-House Calibrations

View

Liquid Handling Knowledge Center

Minimizing Liquid Delivery Risk: Pipetting Technique as Source of Error

View

Liquid Handling Knowledge Center

The Role of Automation in Liquid Delivery Quality Assurance

View