MVS QualAssure® Solutions

Optimize your automated liquid handlers with the ideal liquid for your application

Fast and Straightforward Optimization Process

Optimizing liquid classes can seem tedious but with an MVS, the process is fast and straightforward. Simply set up your assay and use the MVS reagent that best mimics your liquid. You can test different liquid class settings and then use the MVS to measure how the setting affects liquid handling accuracy and precision.

Highlighted features
  • Convenient
  • Easy-to-Use
  • Designed specifically for Artel MVS

Solutions to Mimic your Assay Reagents

When optimizing your automatic liquid handlers (ALHs), it is best to use liquid types similar to what you use your assay. Different liquids have different physical properties (density, viscosity, surface tension), and behave differently when handled by your ALHs. Using water as the default liquid class for your volume verification method can lead to biased data and costly errors down the road. To help you with your ALH optimization, we offer a range of different QualAssure solutions, so you can choose the ones that best mimic your liquid types.

These solutions can measure performance over a wide volume range of 0.0001 to 350 μL and are available for aqueous and non-aqueous volume transfers, including dimethyl sulfoxide (DMSO), PCR master mix-like (PCRMix ™) and serum and plasma-like solutions (SerumSub™ and PlasProxy™)

Liquid Class Optimization for Decreased Variability

MVS QualAssure solutions mimic some of the most common reagents used with liquid handlers, so that you can optimize the performance of your assay, and tailor your liquid handler settings to your specific liquid class, tips and plates. By making the necessary adjustments to your liquid handler, you can ensure accurate and precise liquid handling and minimize the variability of your results.

How It Works:  MVS QualAssure solutions contain two dyes: red and blue. Each solution covers a designated target volume range. As the volume-of-interest decreases, the appropriate QualAssure contains an increased concentration of red dye, thereby maintaining constant signal strength to the plate reader detector. The blue dye is used as an internal standard to measure the solution depth in every well. MVS QualAssure are solutions prepared by Artel and conform to rigorous performance specifications.

Aqueous™ QualAssure®

LEARN MORE

DMSO QualAssure®

LEARN MORE

PCRMix™ QualAssure®

LEARN MORE

SerumSub™ QualAssure®

LEARN MORE

PlasProxy™ QualAssure®

LEARN MORE

Stock Solutions

LEARN MORE

QualAssure Manufacturing

Metrological Traceability

Metrological traceability is maintained to the international system of units (SI). Photometric standards are calibrated at the National Physical Laboratory of the United Kingdom. Qualification of reference spectrophotometers meets applicable standards such as the United States Pharmacopeia (USP), ISO, CLSI, and ASTM International.

Solution Manufacturing

  • Solutions are manufactured in a HEPA filtered clean room.
  • Precise chemical formulations are made according to gravimetric and volumetric standard operating procedures by qualified chemists.
  • Solutions are manufactured, preserved and sub-micron filtered to ensure stability and performance.
  • Release testing is done in Artel’s accredited ISO 17025 laboratory.

Quality Testing

  • Calibration of all solutions is performed with very precise methods.
    • ±0.0004 AU for 730 nm measurements
    • Accuracy measured with uncertainty as low as 0.033%
  • The Artel laboratory maintains an ISO 17025 scope of accreditation for calibration methods used to test all manufactured solutions.
  • All laboratory chemists go through a rigorous training and qualification plan prior to performing methods.
  • Proficiency testing via inter-laboratory and intra-laboratory comparison monitors the competence of all laboratory chemists.
  • The laboratory environment meets the requirements of ISO 8655 and NIST SOP 14.
  • The laboratory maintains a registered quality management system according to ISO 9001 which ensures document control, record control, risk/opportunity, improvement, corrective action, internal audits and management review.

Filling & Labeling

Filling is conducted in a clean room environment. Quality checks are performed throughout the filling and labeling process and records are kept documenting all data collected and indicating personnel involved in each step.

96 Well MVS Verification Plate

Bottle Volume (mL) Solution Range (μL) (Uncertainty) Inaccuracy(%) (Imprecision) CV (%)
Baseline 220 N/A N/A N/A
Diluent 500 N/A N/A N/A
Aqueous HV 500 350.0 to 200.1 1.40 to 1.44 0.15 to 0.16
Aqueous A 500 200.0 to 50.00 1.36 to 1.58 0.16 to 0.19
Aqueous B 110 49.99 to 10.00 1.54 to 1.90 0.15 to 0.20
Aqueous C 110 9.999 to 2.000 1.46 to 1.63 0.15 to 0.20
Aqueous D 110 1.999 to 1.000 1.38 to 1.61 0.16 to 0.19
Aqueous E 110 0.9999 to 0.1000 1.39 to 2.87 0.16 to 0.22
0.0001 to 0.0999 (Non-traceable volume range) N/A N/A
PCRMix B 110 49.99 to 10.00 1.54 to 1.90 0.15 to 0.20
PCRMix C 110 9.999 to 2.000 1.46 to 1.63 0.15 to 0.20
PlasProxy A 500 200.0 to 50.0 3.01 to 3.16 0.16 to 0.19
PlasProxy B 110 49.99 to 10.00 3.00 to 3.16 0.15 to 0.20
DMSO C 110 9.999 to 2.000 1.46 to 1.63 0.15 to 0.20
DMSO D 110 1.999 to 1.000 1.38 to 1.61 0.16 to 0.19
DMSO E 110 0.9999 to 0.1000 1.39 to 2.87 0.16 to 0.22
0.0001 to 0.0999 (Non-traceable volume range) N/A N/A
SerumSub A 500 200.0 to 50.0 3.01 to 3.16 0.16 to 0.19
SerumSub B 110 49.99 to 10.00 3.00 to 3.16 0.15 to 0.20
Stock Solution 1 110 0.4 to 9.9 N/A N/A
Stock Solution 2 110 10 to 49.9 N/A N/A

384 Well MVS Verification Plate

Bottle Volume (mL) Solution Range (μL) (Uncertainty) Inaccuracy(%) (Imprecision) CV (%)
Baseline 220 N/A N/A N/A
Diluent 500 N/A N/A N/A
Aqueous HV 500 N/A N/A N/A
Aqueous A 500 55.00 to 10.00 1.87 to 2.08 0.36 to 0.39
Aqueous B 110 9.999 to 2.500 2.02 to 2.33 0.36 to 0.40
Aqueous C 110 2.499 to 0.500 2.04 to 2.35 0.36 to 0.40
Aqueous D 110 0.4999 to 0.3000 1.88 to 2.01 0.36 to 0.38
Aqueous E 110 0.2999 to 0.0300 1.89 to 3.23 0.36 to 0.44
0.0299 to 0.0100 3.24 to 7.47 0.44 to 0.47
0.0001 to 0.0099 (Non-traceable volume range) N/A N/A
DMSO C 110 2.499 to 0.500 2.04 to 2.35 0.36 to 0.40
DMSO D 110 0.4999 to 0.3000 1.88 to 2.01 0.36 to 0.38
DMSO E 110 0.2999 to 0.0300 1.89 to 3.23 0.36 to 0.44
0.0299 to 0.0100 3.24 to 7.47 0.44 to 0.47
0.0001 to 0.0099 (Non-traceable volume range) N/A N/A
Stock Solution 1 110 0.1 to 2.49 N/A N/A
Stock Solution 2 110 2.5 to 9.9 N/A N/A
MVS® Multichannel Verification System

Simple volume verification with results in less than 10 minutes.

Learn more
ArtelWare™ Software

Ensure your liquid handlers and pipettes are performing properly with full suite of management tools.

Learn more
DMSO QualAssure®

Validates liquid transfers whenever dimethyl sulfoxide solvent is used.

Learn more
SerumSub™ QualAssure®

Acts as a substitute for serum in your most critical assays.

Learn more
PlasProxy™ QualAssure®

Mimics various animal and human plasma in assay verification.

Learn more
PCRMix™ QualAssure®

Mimics commercial PCR master mix in any molecular assays.

Learn more

Request Pricing

For a quotation, to arrange a demonstration, or for more information please contact us by phone or complete the form.

Austria: +49 695 0606 800
Belgium: +31 6 899 10929
France: +33 456 681 861
Germany: +49 695 0606 800
The Netherlands: +31 6 899 10929
United Kingdom: +44 140 321 0400

One moment while the form loads...

Please note, browsing in private or incognito may not be supported and may not allow forms and other elements to load correctly.

Download Brochure

Please fill out this brief form to download this content. You can also get in touch with our sales team for product questions, arranging a product demonstration or obtaining a quotation at the phone numbers below.

Austria: +49 695 0606 800
Belgium: +31 6 899 10929
France: +33 456 681 861
Germany: +49 695 0606 800
The Netherlands: +31 6 899 10929
United Kingdom: +44 140 321 0400

One moment while the form loads...

Please note, browsing in private or incognito may not be supported and may not allow forms and other elements to load correctly.